Skip to content

Public FDA record

Z-2954-2011: IL GEM PCL/PCL Plus - Gem Test PT Cartridge ITC-Nexus Dx A unitized mlcrocoagulation test intended for

Recall record Z-2954-2011 for International Technidyne Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2954-2011
Company in FDA record
International Technidyne Corp.
FDA classification date
Aug 5, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Certain lots of Hemochron Jr. PT Cuvettes (J201) and GEM Test PT cartridges demonstrate a higher bias than historically observed when compared to laboratory reference. instruments.

Source facts

Action
International Technidyne Corp. (ITC) sent "Urgent Medical Device Recall" letters dated May 10, May 25 and June 09, 2011 to all affected customers. The letter included a description of the problem, asked customers fill out the attached form, and return affected product for replacement. For questions, please contact ITC Technical Support at (800) 631-5945.
Initiated
2011-05-10
Root cause
Device Design
Terminated
2012-11-02
Distribution
Worldwide Distribution: (USA) Nationwide Distribution including the countries of; Austria, Australia, Canada, Switzerland, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Korea, Liechtenstien, Libya, Netherlands, Norway, Oman,Portugal, Qatar, Saudi Arabia, Sweden, Turkey, Taiwan, and South Africa.
Recall number
Z-2954-2011
Recall status
Terminated
CDRH record ID
102328
FDA classified
2011-08-05
Recall event number
58043

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗