On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.
Source facts
Action
The firm issued a medical device correction notice on 07/24/2018. The notice explained the issue and provided safety instructions should the use observe a switch change. The units will be corrected. Questions or concerns may contact 1-800-437-1171 or the local service representative.
Initiated
2018-07-24
Root cause
Nonconforming Material/Component
Terminated
2020-12-16
Distribution
Worldwide
Recall number
Z-2954-2018
Recall status
Terminated
CDRH record ID
166164
FDA record created
2018-08-30
Recall event number
80693