Skip to content

Public FDA record

Z-2954-2026: Dentsply Sirona Capsule Mixer Model No. 5546068

Recall record Z-2954-2026 for SDI Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2954-2026
Company in FDA record
SDI Limited
FDA classification date
Aug 14, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for electrical discharge from unit when unplugged.

Source facts

Action
On July 20, 2026, Dentsply Sirona issued a Urgent Medical Device Recall notice to affected consignees via mail. Dentsply Sirona ask consignees to take the following actions: Users 1. Switch off the unit using the rear power switch before removing the plug from the main socket 2. Avoid immediate contact with the plug pins after unplugging. 3. The users should return their unit to Dentsply Sirona to have a corrective modification installed. Please contact Racheal Elliott at rachael.elliott@dentsplysirona.com. 4. Please ensure that all relevant personnel within your organization are informed of the notice. If you have supplied affected units to another organization, please provide this with a copy of the notification. Dealers: 1.Identify all customers who have received the affected units and immediate forward the accompanying communication to those customers. 2.Ensure that all relevant personnel within your organization are informed of this notice. 3.Complete the dealer reply form provided on pages 2 and 3 and rturn it to Dentsply Sirona. 4. Return the completed form via email to Racheal Elliott at rachael.elliott@dentsplysirona.com.
Initiated
2026-07-20
Root cause
Device Design
Distribution
US: AK AL AR AZ CA CO CT DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV OUS: None
Recall number
Z-2954-2026
Recall status
Open, Classified
CDRH record ID
221283
FDA classified
2026-08-14
Recall event number
99322

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗