Open original FDA record ↗Recall record
- FDA identifier
- Z-2955-2011
- Company in FDA record
- St. Jude Medical
- FDA classification date
- Aug 5, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
St. Jude Medical is recalling a single lot (3334664) of Supreme Electrophysiology Extension Cable which were packaged with a Quadripolar cable instead of a hexapolar Cable. St. Jude Medical has determined that the incorrectly labeled product poses no safety risk since the extension cable can be safely used with a corresponding Quadripolar catheter but cannot be incorrectly connected to a Hexapolar catheter.
Source facts
- Action
- St. Jude Medical sent an "IMPORTANT MEDICAL DEVICE FIELD ADVISORY NOTICE" letter dated June 8, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to locate the affected product and to complete and return an attached SJM AFD Reconciliation Form. The form will be used to replace product and as documentation that the customers received the notice. Fax the completed form to 877-497-031. Contact the Director of Product Marketing at 512-329-1054 for questions regarding this notice.
- Initiated
- 2011-06-01
- Root cause
- Labeling mix-ups
- Terminated
- 2012-08-16
- Distribution
- Nationwide Distribution-USA (nationwide) including the states of FL, NC, NY, and OR.
- Recall number
- Z-2955-2011
- Recall status
- Terminated
- CDRH record ID
- 102589
- FDA classified
- 2011-08-05
- Recall event number
- 59392