lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
Source facts
Action
All impacted consignees will be notified by letter delivered by courier service, facsimile or email about the initial notification between August 4th and 6th, 2020. Customers are asked to remove the affected product from use and complete and return the response form to the manufacturer. After response forms are reviewed the customer will be asked to return the product and a replacement will be sent.
Initiated
2020-08-04
Root cause
Mixed-up of materials/components
Terminated
2021-08-12
Distribution
Worldwide distribution - US Nationwide distribution including in the states of MN, TX, CA, UT, WV, FL, IN, GA, AZ, and the countries of Latin America ( CR, MX, TT, EC, PA ), Canada, Europe (FR, AT, BE, ES).
Recall number
Z-2955-2020
Recall status
Terminated
CDRH record ID
183183
FDA classified
2020-09-12
Recall event number
86240