Z-2956-2026: Cardiac Workstation 7000. Product Number: 860441.

Recall record Z-2956-2026 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2956-2026
Company in FDA record
Philips North America Llc
FDA classification date
Aug 14, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device will intermittently provide inaccurate data related to native heart rate.

Source facts

Action
Philips notified consignees on about 07/01/2026 via URGENT Medical Device Correction letter. Consignees were informed that they could continue the use of the device but to refer to the information within the notification letter, disable the automatic Standby feature on the Cardiac Workstation and avoid manually putting the Cardiac Workstation in Standby mode to minimize the inaccurate heart rate value from occurring, and to pass the notice to all those who need to be aware within each organization or to any organization where affected product(s) have been potentially transferred. Additionally, consignees were requested to place the Urgent Medical Device Correction letter with the documentation of the Cardiac Workstation and associated devices in a place where it is most likely to be seen and viewed. Philips representatives will contact consignees to arrange an upgrade to the Cardiac Workstation Release 1.2.0.7.
Initiated
2026-07-01
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Canada, Croatia, Czech Republic, Dominican Republic, France, Germany, Gibraltar, Hong Kong, Iceland, Ireland, Iraq, Italy, Luxembourg, Malta, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom.
Recall number
Z-2956-2026
Recall status
Open, Classified
CDRH record ID
221733
FDA classified
2026-08-14
Recall event number
99361

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗