Open original FDA record ↗Recall record
- FDA identifier
- Z-2957-2026
- Company in FDA record
- Kico Knee Innovation Company
- FDA classification date
- Aug 14, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to an internal sensor issue, image displayed with eyepiece may appear tilted relative to the eyepiece housing
Source facts
- Action
- On 07/13/2026, an "Urgent: Medical Device Recall (Correction)" Letter was email to customers informing them that, KICO has received complaints indicating that the image displayed within the ARVIS eyepiece may appear tilted relative to the eyepiece housing. The display tilt is a persistent condition rather than intermittent or periodic. When the issue occurs, the calibration parameters within an internal sensor change, resulting in a constant offset in the displayed orientation. Once this occurs, the display tilt is consistently present each time the device is powered on and used until the sensor is recalibrated and the firmware correction is implemented. Customers are instructed to: 1. Do not use the ARVIS Gen 2 Navigation System for any hip procedures. 2. The ARVIS Gen 2 Navigation System may continue to be used for Knee and Shoulder procedures only, but it is recommended that all units in inventory are turned on in advance of clinical use and that any units evidencing a tilted display be returned for replacement. 3. This notice needs to be passed on to all those who need to be aware within your organization or functions where the potentially affected devices have been transferred. 4. This notice must be physically posted on or near all ARVIS Gen 2 Navigation Systems. 5. Retain a copy of this letter, place a copy where devices are stored, and keep a copy of the acknowledgement form for your files. 6. Complete the attached Acknowledgement Form immediately and return it email to Alexander.Borchardt@enovis.com 7. The ARVIS Gen 2 Navigation System should always be visually inspected prior to use as stated in the Instructions for Use. 8. Inform KICO of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject devices via the standard complaint process. For question - contact either Sr Product Manager at Jamie.reynolds@enovis.com (512) 451-3815 or Sr Director, Regulatory Affairs at Sherri.Mellingen@enovis.com (952) 913-6383
- Initiated
- 2026-07-13
- Root cause
- Device Design
- Distribution
- Consignee Consignee Number Region Device Quantity RFK MEDICAL CE102 RICHMOND VA IN-29200, ARVIS SURGICAL NAVIGATION 1 VALOR MEDICAL CE118 SAN ANTONIO TX IN-29200, ARVIS SURGICAL NAVIGATION 2 JOINT VENTURES MEDICAL CE121 OSHKOSH WI IN-29200, ARVIS SURGICAL NAVIGATION 1 CASTLE MEDICAL CE122 WATERLOO IA IN-29200, ARVIS SURGICAL NAVIGATION 6 GREAT LAKES MEDICAL GL108 NAPERVILLE IL IN-29200, ARVIS SURGICAL NAVIGATION 2 CERTUS MEDICAL GL112 INDIANAPOLIS IN IN-29200, ARVIS SURGICAL NAVIGATION 5 I&S MEDICAL GL113 LOUISVILLE KY IN-29200, ARVIS SURGICAL NAVIGATION 2 PROCORE MEDICAL MA102 PIEDMONT SC IN-29200, ARVIS SURGICAL NAVIGATION 1 FOX MEDTECH MW001 ENGLEWOOD CO IN-29200, ARVIS SURGICAL NAVIGATION 2 MEDSOURCE NE131 TIVERTON RI IN-29200, ARVIS SURGICAL NAVIGATION 4 IMPLANT MEDICAL SALES NE190 HICKSVILLE NY IN-29200, ARVIS SURGICAL NAVIGATION 2 KINETIC MEDICINE NW105 OMAHA NE IN-29200, ARVIS SURGICAL NAVIGATION 4 CALSURG NW115 SANTA BARBARA CA IN-29200, ARVIS SURGICAL NAVIGATION 4 MIDWEST MEDICAL RESOURCES SC181 OVERLAND PARK KS IN-29200, ARVIS SURGICAL NAVIGATION 3 ORTHO & SURGICAL SOLUTIONS SE175 GUAYNABO PR IN-29200, ARVIS SURGICAL NAVIGATION 3 JOINT MEDICAL SOLUTIONS SE180 BIRMINGHAM AL IN-29200, ARVIS SURGICAL NAVIGATION 4 FLT ORTHOPEDICS SE216 TAMPA FL IN-29200, ARVIS SURGICAL NAVIGATION 6 MAC MEDICAL W180 GILBERT AZ IN-29200, ARVIS SURGICAL NAVIGATION 1 CP ORTHO W214 HONOLULU HI IN-29200, ARVIS SURGICAL NAVIGATION 1 UNIFIED ORTHOPEDICS W223 RIVERSIDE CA IN-29200, ARVIS SURGICAL NAVIGATION 18 MOVEMENT MEDICAL W225 SEATTLE WA IN-29200, ARVIS SURGICAL NAVIGATION 6 Total number of devices in the field 78
- Recall number
- Z-2957-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221438
- FDA classified
- 2026-08-14
- Recall event number
- 99372