Open original FDA record ↗Recall record
- FDA identifier
- Z-2958-2018
- Company in FDA record
- Covidien LLC
- FDA record creation date
- Aug 30, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Revised replacement instructions for the Li-ion battery packs used in BIS Vista and BIS View Monitoring Systems
Source facts
- Action
- Medtronic U.S. consignees were notified via Federal Express or certified mail on July 26, 2018, and the letter informs customers to inspect their batteries. Batteries with date code 3114 or older (Week 31 of 2014) exceed 4 years of age and should be removed from the monitor and disposed of according to facilitys standard process. Included with this notification are stickers to be placed on BIS" monitors when the battery pack is removed the stickers No Battery Installed should be placed on the monitors. The BIS Vista and View monitoring systems may be used without a battery pack only if connected to AC power source with emergency power backup system (Red Outlet).
- Initiated
- 2018-07-26
- Root cause
- Device Design
- Distribution
- Nationwide Foreign: Australia, New Zealand
- Recall number
- Z-2958-2018
- Recall status
- Open, Classified
- CDRH record ID
- 166517
- FDA record created
- 2018-08-30
- Recall event number
- 80804