Open original FDA record ↗Recall record
- FDA identifier
- Z-2959-2026
- Company in FDA record
- Tobii Dynavox Llc
- FDA classification date
- Aug 14, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Source facts
- Action
- On July 8, 2026, the firm notified customers by emailing letters titled "URGENT: MEDICAL DEVICE RECALL NOTICE." Customers were informed that Tobii Dynavox would provide a repair authorization number (RA#), prepaid shipping label, inspection, repair, and return shipping for the affected device. The repair will include modifications intended to improve battery stability within the device. Customers were provided with ways to contact the company to obtain a repair authorization number. Customers were informed that it is safe to continue using the device as intended while waiting for the repair, as long as there are no signs of overheating, swelling, unusual odor, smoke, deformation, or battery abnormality. If any of these conditions are observed, discontinue use immediately and contact Tobbi Dynavox.
- Initiated
- 2026-07-08
- Root cause
- Device Design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2959-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221511
- FDA classified
- 2026-08-14
- Recall event number
- 99410