Z-2960-2026: TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799

Recall record Z-2960-2026 for Tobii Dynavox Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2960-2026
Company in FDA record
Tobii Dynavox Llc
FDA classification date
Aug 14, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

Source facts

Action
On July 8, 2026, the firm notified customers by emailing letters titled "URGENT: MEDICAL DEVICE RECALL NOTICE." Customers were informed that Tobii Dynavox would provide a repair authorization number (RA#), prepaid shipping label, inspection, repair, and return shipping for the affected device. The repair will include modifications intended to improve battery stability within the device. Customers were provided with ways to contact the company to obtain a repair authorization number. Customers were informed that it is safe to continue using the device as intended while waiting for the repair, as long as there are no signs of overheating, swelling, unusual odor, smoke, deformation, or battery abnormality. If any of these conditions are observed, discontinue use immediately and contact Tobbi Dynavox.
Initiated
2026-07-08
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-2960-2026
Recall status
Open, Classified
CDRH record ID
221512
FDA classified
2026-08-14
Recall event number
99410

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗