The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
Source facts
Action
On June 28, 2018, Urgent Medical Device Field Safety Corrective Action letters were distributed to customers. The letters informed the customers of the following:
A Bayer Service Representative will be scheduling a service visit to replace the ceiling column assembly of your OCS system(s) at no cost to you.
Initiated
2018-06-28
Root cause
Equipment maintenance
Terminated
2018-11-26
Distribution
The devices were distributed to the following US states: OH, TX, and WV.
Recall number
Z-2961-2018