Skip to content

Public FDA record

Z-2962-2020: Lactate Dehydrogenase L-P, (LDH) (NAO), Catalog Number LD3842 - Product Usage: device intended for the qu

Recall record Z-2962-2020 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2962-2020
Company in FDA record
Randox Laboratories Ltd.
FDA record creation date
Sep 12, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin. Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.

Source facts

Action
Urgent Medical Device Correction notification letters dated 8/5/20 were issued to customers. Action to be taken: " Review your instrument testing order in line with the updated Carryover Avoidance Technical Bulletin (RXTB-0121) and enable additional pipette washes. " Update the RX user manual with the updated Carryover Avoidance Technical Bulletin (RXTB-0121) and ensure all operators are aware of the recommendations. " Discuss the contents of th is notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. Please accept our apologies for any inconvenience caused. Thank you for your patience and understanding. If you have any questions or concerns, please contact Randox Technical Services.
Initiated
2020-08-11
Root cause
Under Investigation by firm
Terminated
2020-10-15
Distribution
US Nationwide distribution including in the states of WV, and Puerto Rico.
Recall number
Z-2962-2020
Recall status
Terminated
CDRH record ID
183150
FDA record created
2020-09-12
Recall event number
86236

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗