Open original FDA record ↗Recall record
- FDA identifier
- Z-2963-2020
- Company in FDA record
- Randox Laboratories Ltd.
- FDA record creation date
- Sep 12, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin. Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.
Source facts
- Action
- Urgent Medical Device Correction notification letters dated 8/5/20 were issued to customers. Action to be taken: " Review your instrument testing order in line with the updated Carryover Avoidance Technical Bulletin (RXTB-0121) and enable additional pipette washes. " Update the RX user manual with the updated Carryover Avoidance Technical Bulletin (RXTB-0121) and ensure all operators are aware of the recommendations. " Discuss the contents of th is notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. Please accept our apologies for any inconvenience caused. Thank you for your patience and understanding. If you have any questions or concerns, please contact Randox Technical Services.
- Initiated
- 2020-08-11
- Root cause
- Under Investigation by firm
- Terminated
- 2020-10-15
- Distribution
- US Nationwide distribution including in the states of WV, and Puerto Rico.
- Recall number
- Z-2963-2020
- Recall status
- Terminated
- CDRH record ID
- 183151
- FDA record created
- 2020-09-12
- Recall event number
- 86236