Skip to content

Public FDA record

Z-2963-2026: GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventil

Recall record Z-2963-2026 for Maquet Critical Care AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2963-2026
Company in FDA record
Maquet Critical Care AB
FDA classification date
Aug 14, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The module did not undergo electrical safety tests and functional tests during production.

Source facts

Action
Getinge issued an URGENT MEDICAL DEVICE CORRECTION notice was issued to its consignees on 7/9/2026 via FedEx. The notice explained the issue, potential risk to health, and requested the following: "Prior to the replacement, Getinge requests that you take the following actions: 1. Ensure all affected product in your inventory has been identified. 2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 3. Contact Technical Service if you have questions at 888-943-8872 (888 9GETUSA), Option 4, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). 4. Complete and return the enclosed response form to Getinge by e-mailing a copy to recallresponses.qrc@getinge.com. For questions regarding this communication, please contact contact Technical Service if you have any questions at 888-943-8872 (888 9GETUSA), Option 4, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Initiated
2026-06-30
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of CA, GA, MN, NY.
Recall number
Z-2963-2026
Recall status
Open, Classified
CDRH record ID
221556
FDA classified
2026-08-14
Recall event number
99439

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗