A patient record may be linked to an incorrect/inconsistent set of calculation results under specific circumstances when using EyeSuite Software version i2.000.
Source facts
Action
Haag-Streit USA, Inc. sent an "Urgent Field Safety Notices" dated June 20, 2011 to all affected customers.
The letter included a description of the affected product and problem. The firm instructed their customers to make sure all new and current templates are saved and upgrade to EyeSuite version i4.000. Customers are asked to complete and return the Software Update Verification Form.
For additional information contact the Haag-Streit at (513) 398-3937.
Initiated
2011-06-20
Root cause
Software change control
Terminated
2013-03-08
Distribution
Nationwide (USA) Distribution.
Recall number
Z-2964-2011
Recall status
Terminated
CDRH record ID
102758
FDA classified
2011-08-09
Recall event number
59426