Skip to content

Public FDA record

Z-2964-2020: Zenith Alpha Abdominal Endovascular Graft - Distributed OUS only. Indicated for the endovascular treatmen

Recall record Z-2964-2020 for Cook Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2964-2020
Company in FDA record
Cook Inc.
FDA record creation date
Sep 12, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

May contain excess glue within the handle of the delivery system, resulting in an inability to fully deploy the graft via the standard method or troubleshooting deployment method provided in the IFU

Source facts

Action
Cook Medical issued Urgent Medical Device Recall Dated 8/11/20 either mailed via courier or emailed to consignees stating reason for recall, health risk and action to take: Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email FieldActionsNA@CookMedical.com). This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com.
Initiated
2020-08-11
Root cause
Under Investigation by firm
Terminated
2021-07-28
Distribution
Foreign Only: Canada SWEDEN AUSTRIA Belgium CZECH REPUBLIC DENMARK ESPA¿A FINLAND France GERMANY GREECE HUNGARY IRELAND ISRAEL Italia Nederland NORWAY POLAND PORTUGAL REUNION SERBIA SLOVENIA SOUTH AFRICA SWEDEN UNITED ARAB EMIRATES UNITED KINGDOM AUSTRALIA Hong Kong JAPAN NEW CALEDONIA New Zealand Thailand
Recall number
Z-2964-2020
Recall status
Terminated
CDRH record ID
183180
FDA record created
2020-09-12
Recall event number
86242

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗