Skip to content

Public FDA record

Z-2966-2020: AM1054 Randox Ammonia (NH3), Enzymatic UV Method For the quantitative in vitro determination of Ammonia

Recall record Z-2966-2020 for RANDOX LABORATORIES, LTD.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2966-2020
Company in FDA record
RANDOX LABORATORIES, LTD.
FDA record creation date
Sep 14, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Randox Ammonia reagents are being recalled from the field due to a positive bias of up to 140mol/l being observed on patient samples.

Source facts

Action
The firm, Randox Laboratories Ltd., sent an "Urgent Medical Device Correction" letter dated 30th July 2020 to it customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: - Discontinue use of, quarantine and discard any stock of AM1015, batch 517541 and AM1054, batch 512103 immediately. - Review results generated with the affected batches in line with the clinical profile of the patient. - Discuss the contents of this notice with your Medical Director. -Review your reagent inventory of these products and assess your laboratory needs for reimbursement for discarded inventory. -Complete and return the response form 12187-QA to technical.services@randox.com within five working days. If you have any questions or concerns please contact Randox Technical Services. Local at Tel: +1 304 728 2890 Toll Free 866 4 RANDOX Email: customersupportusa@randox.com and PR at Tel: 1-787-701-7000 Email: Jose.Rosario@randox.com.
Initiated
2020-07-30
Root cause
Under Investigation by firm
Terminated
2021-02-03
Distribution
US Distribution to states of: VA, UT, WV and country of : Canada
Recall number
Z-2966-2020
Recall status
Terminated
CDRH record ID
183154
FDA record created
2020-09-14
Recall event number
86238

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗