Z-2966-2026: Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF(s):L2kSY2/10489085; L2KSY6/10488319 IV

Recall record Z-2966-2026 for Siemens Healthcare Diagnostics Products Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2966-2026
Company in FDA record
Siemens Healthcare Diagnostics Products Ltd.
FDA classification date
Aug 14, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient results being reported as nonreactive.

Source facts

Action
On July 13, 2026, Siemens Healthineers issued a Urgent Medical Device Correction to affected consignees via FedEx. Siemens asked consignees to take the following actions: 1. Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2.Discontinue use and discard the affected kit lot. 3. Review your current inventory to determine your laboratory s replacement needs and provide this information to Siemens Healthineers for reporting to the authorities. 4. Complete and return the Field Correction Effectiveness Check Form attached to this letter within (30) days. 5.Retain this letter with your laboratory records and forward this letter to those who may have received the affected product(s).
Initiated
2026-07-13
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of FL, NJ, NY and the countries of Austria, Azerbaijan, Belgium, Brazil, Chile, Colombia, Cura-ao, St.¿Eus Cyprus, France, Germany, Italy, Kazakhstan, Latvia, Lithuania, Netherlands, Romania, Russian¿Fed., Saudi¿Arabia, Slovakia, Slovenia, South¿Africa, Spain, U.A.E., Ukraine, United¿Kingdom, Uzbekistan.¿¿¿¿¿
Recall number
Z-2966-2026
Recall status
Open, Classified
CDRH record ID
221576
FDA classified
2026-08-14
Recall event number
99447

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗