Z-2967-2020: Tosoh Bioscience AIA- PACK HCG Calibrator Set-In Vitro Diagnostic for BHCG Part number 020361 The AIA-Pa

Recall record Z-2967-2020 for Tosoh Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2967-2020
Company in FDA record
Tosoh Bioscience Inc
FDA classification date
Sep 14, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.

Source facts

Action
The firm, Tosoh Bioscience, issued an "URGENT: MEDICAL DEVICE RECALL" Letter dated August 7, 2020 via Fedex to its customers. The letter states reason for recall, health risk and action to taken: Check the lot numbers of the AIA - PACK HCG Calibrator Set and AIA-PACK HCG Calibration Verification Test Set against the lot numbers listed above. Lots not listed above are not within the scope of this recall and maybe used per AAM. Remove all affected AIA - PACK HCG Calibrator Sets and AIA - PACK HCG Calibration Verification Test Sets from your inventory. The products must be tallied including previously used sets of the affected lots and properly destroyed per your local procedures and regulations. For reimbursement of destroyed calibrator and linearity sets, please contact Technical Support. Reimbursement will be with same or similar products. Review the information in this notification with the Medical Director and/or Lab Director as soon as possible and review the content of this letter with your laboratory staff Complete and return the attached Acknowledgement Form to Tosoh Bioscience, Inc., within 15-days of receiving this notification via Fax: 1-650-636-8651 or Email: biorecallresponse@tosoh.com or by US Post Mail: Be sure to indicate destruction of affected product(s). Tosoh Bioscience Inc. 3600 Gantz Rd. Grove City, OH 4312. Maintain this notification with your laboratory records and forward this information to others in your laboratory. If you have obtained unexpected test results or received any complaints of illness or adverse events associated with the use of the analyzer, please contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. Should you have any questions regarding this medical device recall, please feel free to contact Bernadette OConnell at (800) 248-6764 or by email at bernadette.oconnell@tosoh.com. Monday - Friday, from 9:00 AM to 5:00 PM (PST).
Initiated
2020-08-07
Root cause
Under Investigation by firm
Terminated
2021-06-21
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Bogota, Cayman Islands, Mexico, Peru, and San Salvador.
Recall number
Z-2967-2020
Recall status
Terminated
CDRH record ID
183112
FDA classified
2020-09-14
Recall event number
86218

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗