Skip to content

Public FDA record

Z-2967-2026: Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001

Recall record Z-2967-2026 for Paragonix Technologies Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2967-2026
Company in FDA record
Paragonix Technologies Inc.
FDA classification date
Aug 14, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.

Source facts

Action
On July 20, 2026, Paragonix Technologies Inc./Getinge issued an Urgent Medical Device Correction to affected consignees via FedEX mail. Paragonix requested consignees to take the following actions: 1. An update to the Paragonix App has been released through the application store. 2. We recommend updating all mobile devices to latest version prior to KidneyVault usage. 3. Please continue to inspect all KidneyVault devices per the IFU before use to ensure proper datalogger function. 4. Continue to follow the IFU for using a multi-faceted clinical approach to assessing the organ, including any/all information available on the datalogger screen. 5. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 6. Complete and return the enclosed response form by e-mailing a copy to recallresponses.qrc@getinge.com.
Initiated
2026-07-20
Root cause
Software design
Distribution
US Nationwide distribution in the states and territories of AL, AZ, CA, TN, FL, HI, MN, NJ, TX, UT, CO, CT, DC, MD, GA, KY, LA, MA, MI, MO, MS, NC, ND, NE, NM, NV, NY, PA, PR, SC, UT, VA, WI, WV, IN, OH, KY, IL, MD, MN, OR.
Recall number
Z-2967-2026
Recall status
Open, Classified
CDRH record ID
221632
FDA classified
2026-08-14
Recall event number
99476

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.