Open original FDA record ↗Recall record
- FDA identifier
- Z-2968-2018
- Company in FDA record
- McKesson Israel Ltd.
- FDA record creation date
- Aug 31, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Change Healthcare has identified an issue where, under certain circumstances, the Real Time Monitor (RTM) may not display physiological signals.
Source facts
- Action
- The firm, Change Healthcare Israel Ltd., sent an "Urgent Field Safety Notice" dated on 03/12/2018 to its customers. The notice explained the product, issue and actions to be taken. The customers were instructed to: distribute the notice to all those within your organization who need to be aware; alert other parties affiliated with your organization, which are affected by this field safety notice; and maintain awareness of this notice and resulting action until the defect has been corrected, to ensure effectiveness of the corrective action. Change Healthcare has developed software updates to address the problem. If you have any questions, contact Change Healthcare Customer Support (US) 1-800-654-4366 or (UK) 440(0) 208-952-799 to arrange for the installation of software update/upgrade.
- Initiated
- 2018-03-12
- Root cause
- Software Design Change
- Terminated
- 2021-11-23
- Distribution
- US Distribution to states of: CA, CO, CT, FL, GA, KS, IN, LA, NJ, NY, OK, OH, SC, and TX; and internationally to: UK.
- Recall number
- Z-2968-2018
- Recall status
- Terminated
- CDRH record ID
- 166225
- FDA record created
- 2018-08-31
- Recall event number
- 80723