Open original FDA record ↗Recall record
- FDA identifier
- Z-2969-2026
- Company in FDA record
- Gentherm Medical, LLC
- FDA classification date
- Aug 14, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Overheating of the 115V CPU board may cause the device to shut down.
Source facts
- Action
- Genthem issued an URGENT Medical Device Notice to its consignees on 7/17/2026 via email. The notice explained the issue, potential risk, and request the following: "Interim Recommendations: 1. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available. 2. Users should review and follow the device Instructions for Use; users should routinely verify the device s function and patient s condition during use. 3. Ensure that all users are informed of the contents of this letter. If you have further distributed this product, please provide those accounts with a copy of this notice." For questions or if you have experienced the described issue or a related adverse event, call Gentherm Medical Technical Support at 1-888-437-5608.
- Initiated
- 2026-07-13
- Root cause
- Device Design
- Distribution
- Worldwide - US Nationwide distribution including in the territories of Puerto Rico, and the countries of Brazil, Canada, Chile, Colombia, Costa Rica, El Salvador, Guatemala, Mexico, Republic of Panama, Saudi Arabia, Taiwan, Thailand, Bahamas, and Trinidad and Tobago.
- Recall number
- Z-2969-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221635
- FDA classified
- 2026-08-14
- Recall event number
- 99478