Skip to content

Public FDA record

Z-2970-2018: DORO LUCENT Base Unit, REF 1101.021 Headrest System used in open and percutaneous craniotomies and spina

Recall record Z-2970-2018 for Pro-Med Instruments Gmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2970-2018
Company in FDA record
Pro-Med Instruments Gmbh
FDA classification date
Sep 17, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.

Source facts

Action
The firm initiated the recall by email on 07/05/2018. This stage is only relevant for hospitals / users who are in possession with concerned products of two different revisions (mix-up): a) Discontinue the use of the combination and quarantine the concerned products b) Contact pro med instruments for further instructions relating to temporary replacement possibilities
Initiated
2018-07-05
Root cause
Component change control
Terminated
2020-11-19
Distribution
Worldwide distribution to US states of CA, TX, TN, MS, GA, WV, and OH, Qatar, India, UK, Singapore, Australia, Germany, South Africa, Italy, and Japan
Recall number
Z-2970-2018
Recall status
Terminated
CDRH record ID
165915
FDA classified
2018-09-17
Recall event number
80628

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗