Skip to content

Public FDA record

Z-2970-2020: Siemens epoc BGEM Test Card-In vitro diagnostic device for the quantitative testing of samples of heparin

Recall record Z-2970-2020 for Siemens Healthcare Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2970-2020
Company in FDA record
Siemens Healthcare Diagnostics Inc
FDA record creation date
Sep 15, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Sporadically inconsistent discrepant (low bias) glucose results on card lot 01-20095-10. Potential exists for a delay in diagnosis of hyperglycemia or unnecessary treatment for hypoglycemia.

Source facts

Action
Siemens issued Urgent Medical Device Correction letter on August 28, 2020 via Federal Express stating the reason for recall communication letter informs customers of the issue and, health risk and action take: Please dispose any inventory of test card lot 01-20095-10 currently in your possession in accordance with local and state disposal requirements. You may request free of charge replacement product from your local Siemens or distributor office. Please review your inventory of these products and assess your laboratorys replacement needs. Please use epoc BGEM Test Card from another unaffected lot. Complete and return the Field Correction Effectiveness Check and Product Replacement Form attached to this letter. Please retain this letter with your laboratory records and forward it to those who may have received this product. Questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative.
Initiated
2020-08-28
Root cause
Process control
Terminated
2021-09-09
Distribution
Nationwide Foreign: Canada
Recall number
Z-2970-2020
Recall status
Terminated
CDRH record ID
183514
FDA record created
2020-09-15
Recall event number
86378

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.