Skip to content

Public FDA record

Z-2970-2026: Blanketrol III, Model 233; Thermal regulating system.

Recall record Z-2970-2026 for Gentherm Medical, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2970-2026
Company in FDA record
Gentherm Medical, LLC
FDA classification date
Aug 14, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Overheating of the 115V CPU board may cause the device to shut down.

Source facts

Action
Genthem issued an URGENT Medical Device Notice to its consignees on 7/17/2026 via email. The notice explained the issue, potential risk, and request the following: "Interim Recommendations: 1. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available. 2. Users should review and follow the device Instructions for Use; users should routinely verify the device s function and patient s condition during use. 3. Ensure that all users are informed of the contents of this letter. If you have further distributed this product, please provide those accounts with a copy of this notice." For questions or if you have experienced the described issue or a related adverse event, call Gentherm Medical Technical Support at 1-888-437-5608.
Initiated
2026-07-13
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution including in the territories of Puerto Rico, and the countries of Brazil, Canada, Chile, Colombia, Costa Rica, El Salvador, Guatemala, Mexico, Republic of Panama, Saudi Arabia, Taiwan, Thailand, Bahamas, and Trinidad and Tobago.
Recall number
Z-2970-2026
Recall status
Open, Classified
CDRH record ID
221636
FDA classified
2026-08-14
Recall event number
99478

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗