Open original FDA record ↗Recall record
- FDA identifier
- Z-2971-2011
- Company in FDA record
- AMO Manufacturing USA, LLC
- FDA classification date
- Aug 9, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Reports of lower success rates with Iris Registration (IR) image capture with the STAR S4 IR Excimer Laser System v 5.3 that with previous version.
Source facts
- Action
- Abbott Medical Optics (AMO) sent an "ADVISORY NOTICE" dated July 18, 2011 to all customers using the AMO WAVESCAN Wavefront System. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to create a separate patient file for each eye on the WAVESCAN System (STAR S4 IR Excimer Laser Systems,software versions 5.00-5.30) according to the directions in attachment A. Additionally, customers are to complete and return an attached form to AMO via fax at the regional fax number. Contact an AMO representative for questions regarding this notice.
- Initiated
- 2011-07-18
- Root cause
- Software design
- Terminated
- 2013-09-10
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico and countries of Africa, Asian Pacific, Canada, Europe, Japan, Latin America, and the Middle East.
- Recall number
- Z-2971-2011
- Recall status
- Terminated
- CDRH record ID
- 102306
- FDA classified
- 2011-08-09
- Recall event number
- 59330