Skip to content

Public FDA record

Z-2971-2011: STAR S4 IR Excimer Laser System; Software Version 5.18 Mfg by : AMO Manufacturing USA, LLC 510 Cottonwo

Recall record Z-2971-2011 for AMO Manufacturing USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2971-2011
Company in FDA record
AMO Manufacturing USA, LLC
FDA classification date
Aug 9, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Reports of lower success rates with Iris Registration (IR) image capture with the STAR S4 IR Excimer Laser System v 5.3 that with previous version.

Source facts

Action
Abbott Medical Optics (AMO) sent an "ADVISORY NOTICE" dated July 18, 2011 to all customers using the AMO WAVESCAN Wavefront System. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to create a separate patient file for each eye on the WAVESCAN System (STAR S4 IR Excimer Laser Systems,software versions 5.00-5.30) according to the directions in attachment A. Additionally, customers are to complete and return an attached form to AMO via fax at the regional fax number. Contact an AMO representative for questions regarding this notice.
Initiated
2011-07-18
Root cause
Software design
Terminated
2013-09-10
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and countries of Africa, Asian Pacific, Canada, Europe, Japan, Latin America, and the Middle East.
Recall number
Z-2971-2011
Recall status
Terminated
CDRH record ID
102306
FDA classified
2011-08-09
Recall event number
59330

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗