Open original FDA record ↗Recall record
- FDA identifier
- Z-2972-2020
- Company in FDA record
- Imactis
- FDA record creation date
- Sep 15, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiducial could lead to inability to navigate or inaccuracies such as improper needle trajectory.
Source facts
- Action
- On 08/XX/2020, affected consignees where emailed a copy of the "Urgent - Medical Device Correction" notification. In addition, a paper copy will be provided to each consignee in person by the end of August 2020. In addition to providing information on the correction, the notification asked consignees to take the following actions: 1. Inspect your patient fiducial for potential separation and confirm to us, using the attached acknowledgement of receipt (See Exhibit A). - The fiducial disk is normally flush with the surface. Check the fiducial disk to determine that it is at the same level as the fiducial body. If the fiducial disk is raised from the body surface, then there is a separation. 2. If disk separation is noticed: - Remove the CT-Navigation" workstation from service. - Immediately contact your IMACTIS representative for repair or replacement - Robert Hornak - 205-542-1486 (cell), Monday through Friday, 8:00 AM to 6:00 PM, Eastern Time 3. A warning label will be sent to your facility and must be affixed to the IMACTIS CT-Navigation" workstation (See attached Exhibit B). 4. Complete the Exhibit A "Acknowledgement of Receipt of Important Customer Information" and email to quality@imactis.com 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
- Initiated
- 2020-08-24
- Root cause
- Use error
- Distribution
- US Nationwide distribution including in the states of FL & WI.
- Recall number
- Z-2972-2020
- Recall status
- Open, Classified
- CDRH record ID
- 180372
- FDA record created
- 2020-09-15
- Recall event number
- 85258