Skip to content

Public FDA record

Z-2972-2020: Imactis Patient Fiducial (Patient Referential), REF: IHR-YY-XXX component used in conjunction with the

Recall record Z-2972-2020 for Imactis. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2972-2020
Company in FDA record
Imactis
FDA record creation date
Sep 15, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiducial could lead to inability to navigate or inaccuracies such as improper needle trajectory.

Source facts

Action
On 08/XX/2020, affected consignees where emailed a copy of the "Urgent - Medical Device Correction" notification. In addition, a paper copy will be provided to each consignee in person by the end of August 2020. In addition to providing information on the correction, the notification asked consignees to take the following actions: 1. Inspect your patient fiducial for potential separation and confirm to us, using the attached acknowledgement of receipt (See Exhibit A). - The fiducial disk is normally flush with the surface. Check the fiducial disk to determine that it is at the same level as the fiducial body. If the fiducial disk is raised from the body surface, then there is a separation. 2. If disk separation is noticed: - Remove the CT-Navigation" workstation from service. - Immediately contact your IMACTIS representative for repair or replacement - Robert Hornak - 205-542-1486 (cell), Monday through Friday, 8:00 AM to 6:00 PM, Eastern Time 3. A warning label will be sent to your facility and must be affixed to the IMACTIS CT-Navigation" workstation (See attached Exhibit B). 4. Complete the Exhibit A "Acknowledgement of Receipt of Important Customer Information" and email to quality@imactis.com 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Initiated
2020-08-24
Root cause
Use error
Distribution
US Nationwide distribution including in the states of FL & WI.
Recall number
Z-2972-2020
Recall status
Open, Classified
CDRH record ID
180372
FDA record created
2020-09-15
Recall event number
85258

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗