Z-2973-2020: FlexLab (FLX) with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Trac

Recall record Z-2973-2020 for Inpeco S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2973-2020
Company in FDA record
Inpeco S.A.
FDA record creation date
Sep 16, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results

Source facts

Action
Inpeco issued Customer Letter dated August 7, 2020 forwarded to the customers either by Inpeco (for the laboratories where Inpeco is directly the distributor) or by the Inpeco distributors. Letter states reason for recall, health risk and action to take: customers impacted by the Field Action have to complete and return the Recall Letter Receipt Confirmation and Implementation Check form attached to the Customer letter within 15 days to Inpeco or the distributor. The customers impacted by the issue will be contacted by Inpeco or the distributors to arrange a visit to upgrade the impacted modules firmware. Contact reference person: Eva Balzarotti - Regulatory Affairs Manager. E-mail: Regulatory.Affairs@inpeco.com Phone: (+41) 919118 224
Initiated
2020-08-07
Root cause
Software design
Terminated
2021-12-02
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, FL, IA, LA, MA, MS, MT, NC, OH, TX, VA and the countries of AUSTRALIA, AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, NETHERLANDS. NORWAY SAUDI ARABIA SOUTH KOREA SWEDEN UNITED ARAB EMIRATES
Recall number
Z-2973-2020
Recall status
Terminated
CDRH record ID
183241
FDA record created
2020-09-16
Recall event number
86263

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗