Z-2975-2024: 25-LHP-828, HeNe Laser System, 35 mW

Recall record Z-2975-2024 for PACIFIC LASERTECH, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2975-2024
Company in FDA record
PACIFIC LASERTECH, LLC
FDA classification date
Sep 10, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.

Source facts

Action
The firm notified all purchasers of the defect. Customers were instructed to return all units for replacement of the switch free of charge.
Initiated
2024-03-01
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide Distribution
Recall number
Z-2975-2024
Recall status
Open, Classified
CDRH record ID
209910
FDA classified
2024-09-10
Recall event number
95290

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗