These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.
Source facts
Action
The firm notified all purchasers of the defect. Customers were instructed to return all units for replacement of the switch free of charge.
Initiated
2024-03-01
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide Distribution
Recall number
Z-2976-2024
Recall status
Open, Classified
CDRH record ID
209911
FDA classified
2024-09-10
Recall event number
95290
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.