Open original FDA record ↗Recall record
- FDA identifier
- Z-2977-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Aug 10, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A potential safety rick related to transferring reference images for offset calculation between a syngo RT Therapist v 4.2 or a syngo RT Oncologist 4.2 and other versions of the COHERENCE/syn go RT Therapist or Oncologist or vice versa. An incorrect offset may be applied for patient positioning and result in dose being delivered to the wrong location.
Source facts
- Action
- Siemens sent an "URGENT: MEDICAL DEVICE CORRECTION/CUSTOMER SAFETY ADVISORY NOTICE" letter dated June 3, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter provides preventive measure instructions for customers to use until an update for the system is available.
- Initiated
- 2011-06-03
- Root cause
- Software design
- Terminated
- 2012-03-30
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of CA, FL, NC, NY, OH, PA, SD, UT, and WI and countries of Australia, Belgium, Croatia, Czech Republic, Germany, India, Ireland, Japan, Poland, Saudi Arabia, South Africa, Spain, Sweden, and United Kingdom.
- Recall number
- Z-2977-2011
- Recall status
- Terminated
- CDRH record ID
- 101206
- FDA classified
- 2011-08-10
- Recall event number
- 59107