Skip to content

Public FDA record

Z-2978-2018: Laparoscopy CDS

Recall record Z-2978-2018 for Medline Industries, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2978-2018
Company in FDA record
Medline Industries, Inc.
FDA record creation date
Sep 6, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

One manufacturing lot of a non-sterile pack was assembled with a non-sterile component labeled sterile (kit component #80532NS). The non-sterile component 80532NS was supposed to be placed inside the sterile portion of the pack where it would subsequently be exposed to the ethylene oxide sterilization process. However, the component was not placed in the sterile portion of the pack during assembly and was never exposed to the sterilization process. Component 80532NS was inadvertently released non-sterile. Component 80532NS is packaged by Medline and is an Anti-fog kit that contains 1 bottle of 6g Anti-Fog solution and an adhesive backed sponge.

Source facts

Action
Firm contacted customers via phone and letter beginning April 12, 2017. Customers were instructed to remove the Ethicon BioPatch Protective Disk with CHG from the kit(s) and return it to the firm. The firm would provide a sterile replacement BioPatch. Affected product will be destroyed.
Initiated
2017-04-12
Root cause
Under Investigation by firm
Terminated
2019-08-05
Distribution
CA, NM
Recall number
Z-2978-2018
Recall status
Terminated
CDRH record ID
159474
FDA record created
2018-09-06
Recall event number
78411

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗