Skip to content

Public FDA record

Z-2978-2020: The TaqPath RT-PCR COVID-19 Kit with Applied Biosystem's COVID-19 Interpretive Software v1.2 for 7500 RTP

Recall record Z-2978-2020 for Life Technologies Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2978-2020
Company in FDA record
Life Technologies Corporation
FDA record creation date
Sep 17, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

COVID-19 Interpretive Software issues that may cause: 1) the risk of invalid or inconclusive tests/false negative and potentially increasing the retesting burden on customers. 2) false positive results due to improper vortexing.

Source facts

Action
On 08/06/2020, the firm sent an "Urgent: Medical Device Field Correction" notification to customers via email and a follow-up letter via FedEx informing them of two issues associated with COVID-19 interpretive software: 1) A small number of poorly extracted patient samples were called valid by the interpretive software, related to current MS2 assay analysis settings. This issue can potentially cause a weakly positive sample that is prepared by suboptimal extraction to be falsely called a valid negative sample, thereby constituting a False Negative. And 2) The MS2 assay, which detects the Internal Positive Control (IPC), was erroneously called amplified in a small percentage of Positive Control (PC) samples. This issue caused a plate to be designated invalid, thereby requiring unnecessary retesting of an entire batch of samples. The firm is not requesting return of product. The issue is related to software and improper mixing/vortexing of the qPCR plates. Customer are instructed to: -Notify all Applied Biosystems COVID-19 Interpretive Software users in their facility of this communication. -Stop using Applied Biosystems COVID-19 Interpretive Software v1.2, v2.0 and v2.2. -Go to www.thermofisher.com/educationconnect and sign in with their thermofisher.com username and password. Type in the subscription code C19CKEUAFIS1323 to access and complete the training. The customer will be required to pass an exam and acknowledge that they have reviewed information to upgrade the COVID-19 Interpretive Software. Re-registration is required for this upgrade. For full instructions to learn how to access the new version of the software please read the recalling firm's EducationConnect welcome letter or watch the quick video tutorial. -Read and implement the vortexing recommendations added to the latest release of the Instructions for Use available on the product page TaqPath COVID-19 EUA IFU to prevent insufficient mixing and/or centrifugation. -Sign and return Ac
Initiated
2020-08-06
Root cause
Other
Terminated
2022-05-05
Distribution
US: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NM, NY, NV, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. OUS: Afghanistan, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Botswana, Brazil, Cameroon, Canada, Cypress, Estonia, France, Georgia, Germany, Greece, Guadeloupe, Hungary, India, Indonesia, Ireland, Israel, Italy, Kenya, Kuwait, Lebanon, Mozambique, Namibia, New Zealand, Oman, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, United Republic of Tanzania, and Vietnam.
Recall number
Z-2978-2020
Recall status
Terminated
CDRH record ID
183031
FDA record created
2020-09-17
Recall event number
86204

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗