Open original FDA record ↗Recall record
- FDA identifier
- Z-2979-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Aug 10, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Firm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing necessary.The potential exists for a cable connection to become overheated, leading to the emission of smoke and possible material damage.
Source facts
- Action
- Siemens sent a "CUSTOMER SAFETY ADVISORY NOTICE" letter dated June 20, 2011 to all affected customers. The letter describes the product, problem, and actions that will be taken. The letter informs customers that an inspection of the gradient cable connections and an emergency shutdown mechanism will be installed by a field service representative during a visit to all affected locations.
- Initiated
- 2011-06-20
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-01-14
- Distribution
- Nationwide Distribution-USA (nationwide) including the states of AK, FL, GA, IL, MA, MI, MN, MT, NC, NJ, NY, OH, TX, UT, VA, WA, and WV.
- Recall number
- Z-2979-2011
- Recall status
- Terminated
- CDRH record ID
- 101482
- FDA classified
- 2011-08-10
- Recall event number
- 59170