Open original FDA record ↗Recall record
- FDA identifier
- Z-2980-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Aug 10, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There a potential patient safety issue when using the Symbia Radiation Therapy Pallet. When a patient is not iso-centered before starting a scan with the Symbia T Series Systems and a non-circular orbit is selected, it is possible that one of the detectors may come into contact with the patient.
Source facts
- Action
- Siemens Medical Solutions USA, Inc. sent an "Urgent Field Correction Recall" letter dated June 06, 2011 to all affected customers. The letter informs customers of the potential safety issue if the patient is not iso-centered before starting a scan and a non-circular orbit is selected and provided recommendations including ensure the patient is iso-centered before starting a scan. If the patient cannot be positioned at the iso-center of the RTP, instructions were provided. The customers are asked to complete the Mandatory Fax-Back Form acknowledging receipt and understanding of the letter. For questions or assistance contact Siemens Medical Solutions USA, Inc. at (800) 888-7436 (USA).
- Initiated
- 2011-06-22
- Root cause
- Device Design
- Terminated
- 2012-11-27
- Distribution
- Worldwide Distribution: Nationwide (USA) including the states of Florida and Texas; and the countries of Canada, China, Italy, Mexico, Poland, Portugal and Russia.
- Recall number
- Z-2980-2011
- Recall status
- Terminated
- CDRH record ID
- 100609
- FDA classified
- 2011-08-10
- Recall event number
- 58918