Z-2980-2020: The Guardian Connect App CSS7200 iOS - Product Usage: intended for use by patients with a compatible cons

Recall record Z-2980-2020 for Medtronic Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2980-2020
Company in FDA record
Medtronic Inc.
FDA record creation date
Sep 17, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

As a result of the release of new software version to CareLink Personal website, the IOS app for the Continuous Monitoring Glucose system stopped automatically uploading data to the website.

Source facts

Action
On July 15, 2020, the firm sent an "URGENT FIELD SAFETY NOTIFICATION" to customers/Care Partners/Healthcare Professionals via email informing them that the Recalling Firm released an update to the CareLink Personal Software. Following the release of the software, several customer using the iOS app experienced data connectivity disruptions and as a result: 1. Care partners may not receive SMS alerts 2. Care partners remote access monitoring in the CareLink Connect web app may show data gaps 3. Customers may see data gaps in their reports. Customer are instructed to: 1. Attempt to manually upload your data to the CareLink" software via the Guardian" Connect app: a. Tap on the top left corner of the Home Screen b. Confirm Sync to CareLink is enabled c. Tap Upload Now d. Wait 10 minutes before moving to step #2. 2. Check your account to see if you are experiencing this issue a. From your mobile device visit https://carelink.minimed.eu and log in to your CareLink" account b. Confirm that your data is visible in the CareLink" Connect web app c. If your data is visible, no further action is needed - you are not experiencing the issue d. If your data is not visible, you may be experiencing the issue and you should complete the steps in the next section to resolve the issue. Instructions for Resolving the Issue (refer to the Frequently Asked Questions included for detailed instructions) 1. If your data was not visible after step #2, uninstall the Guardian" Connect app from your mobile device. For help, please refer to FAQ below #3 2. Unpair your transmitter from your mobile device. For help, please refer to FAQ below #3 3. Re-install the Guardian" Connect app on your mobile device this will restore the CareLink" connection. For help, please refer to FAQ below #4 *Uninstalling the Guardian" Connect app may delete visible data from the app. All data that has been uploaded to the CareLink" software will continue to be visible in CareLink" report
Initiated
2020-07-15
Root cause
Software design
Terminated
2022-02-18
Distribution
International distribution including in the countries of EMEA - Austria, Belgium, Croatia, Czechia, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Kuwait, Latvia, Luxembourg, The Netherlands, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, and United Kingdom. Latin America- Chile, Colombia and Mexico Asia Pacific (APAC) - Australia, India, Japan, The Republic of Korea, Malaysia, New Zealand, Russia Federation, Singapore, Thailand, and Vietnam. Asia - Hong Kong North America - Canada
Recall number
Z-2980-2020
Recall status
Terminated
CDRH record ID
182620
FDA record created
2020-09-17
Recall event number
86081

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗