Skip to content

Public FDA record

Z-2981-2020: Elecsys Anti-TPO Assay on cobas e analyzers 411, 601, 602, Catalog Number 06368590190

Recall record Z-2981-2020 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2981-2020
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA record creation date
Sep 17, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Recent internal verification studies found the current RF interference claim to no longer be met for the assay on the cobas e 411, 601, and 602 analyzers. Therefore, the claim has been updated.

Source facts

Action
Urgent Medical Device Correction notification letters dated 8/20/20 were distributed to customers. Actions Required " Be aware of the new RF interference claims listed in this Urgent Medical Device Correction (UMDC). This UMDC serves as labeling until the updated method sheet (V3), anticipated in September 2020, is available. " Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. Customers do not need to perform an additional RF test unless otherwise determined in consultation with the physician or pathologist at your facility. " Complete the attached fax back form (TP-01048) and fax or email it according to the instructions on the form. " File this UMDC for future reference. Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. The Elecsys Anti-TPO method sheet is being updated with the new claim for Rheumatoid Factor (RF) interference. Until the updated method sheet (V3) is available, this Urgent Medical Device Correction serves as your labeling. When these method sheets are updated, they will be available on the diagnostics.roche.com website and on the cobas e-library. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Initiated
2020-08-20
Root cause
Nonconforming Material/Component
Terminated
2021-10-22
Distribution
US Nationwide
Recall number
Z-2981-2020
Recall status
Terminated
CDRH record ID
183335
FDA record created
2020-09-17
Recall event number
86319

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗