Skip to content

Public FDA record

Z-2982-2011: Biopsy Trays (CT, Puncture & Prep), Packaged individually in header bags, 10 per case. Sterile, Disposab

Recall record Z-2982-2011 for North Coast Medi-Tek Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2982-2011
Company in FDA record
North Coast Medi-Tek Inc
FDA classification date
Aug 10, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During routine bioburden testing, unusually high results were found for a Bioburden Test done on 4/19/11. Bioburden Level reported as >150,000 CFU. Our Alert Level is 17,000 CFU and the Action Level is 23,000 CFU.

Source facts

Action
North Coast Medi - Tek Inc. sent a URGENT PRODUCT RECALL letter dated May 13, 2011 to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. The customers were instructed to stop shipment of all remaining product and return to recalling firm. They were also requested to contact their customers and notify them of the recall and retieve all remaining product. For further questions please call (440) 974-0750.
Initiated
2011-05-13
Root cause
Other
Terminated
2012-02-16
Distribution
Nationwide Distribution CA, CO, FL, GA, ID, IL, MA, NC, NE, NY, OH, OK, PA, TN, NC and TX
Recall number
Z-2982-2011
Recall status
Terminated
CDRH record ID
102584
FDA classified
2011-08-10
Recall event number
59386

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗