During routine bioburden testing, unusually high results were found for a Bioburden Test done on 4/19/11. Bioburden Level reported as >150,000 CFU. Our Alert Level is 17,000 CFU and the Action Level is 23,000 CFU.
Source facts
Action
North Coast Medi - Tek Inc. sent a URGENT PRODUCT RECALL letter dated May 13, 2011 to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer.
The customers were instructed to stop shipment of all remaining product and return to recalling firm. They were also requested to contact their customers and notify them of the recall and retieve all remaining product.
For further questions please call (440) 974-0750.
Initiated
2011-05-13
Root cause
Other
Terminated
2012-02-16
Distribution
Nationwide Distribution CA, CO, FL, GA, ID, IL, MA, NC, NE, NY, OH, OK, PA, TN, NC and TX
Recall number
Z-2982-2011
Recall status
Terminated
CDRH record ID
102584
FDA classified
2011-08-10
Recall event number
59386