Open original FDA record ↗Recall record
- FDA identifier
- Z-2982-2026
- Company in FDA record
- Medtronic MiniMed, Inc.
- FDA classification date
- Aug 18, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Source facts
- Action
- On 7/13/2026, field safety notices were sent to customers asking them to do the following: 1) Stop using the loaner pump and contact the firm to arrange for loaner pump return. 2) Contact your Healthcare Professional (HCP) to ensure you have access to backup insulin therapy supplies or alternative plans, which may include short- and long-acting insulins, syringes, with dosage instructions from your HCP. 3) Complete and return the customer acknowledgement form. Questions can be directed to the firm's Tech Support: 1-800-646-4633
- Initiated
- 2026-07-13
- Root cause
- Process design
- Distribution
- International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
- Recall number
- Z-2982-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221570
- FDA classified
- 2026-08-18
- Recall event number
- 99374