Open original FDA record ↗Recall record
- FDA identifier
- Z-2983-2020
- Company in FDA record
- Beckman Coulter Biomedical GmbH
- FDA record creation date
- Sep 18, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The DxA 5000 instructions for use may be confusing regarding the re-use of sample IDs. Beckman Coulter has received complaints where the re-use of sample barcodes were not being detected by the system .
Source facts
- Action
- The Urgent Medical Device Recall letters were sent via first class mail on 06/19/20. The actions to be taken by consignees are as follows: Avoid Duplicate IDs; DO NOT Re-use Sample IDs As indicated in the DxA 5000 instructions for use, ensure that only unique sample barcode IDs are used. Be aware of the difference between duplicate sample IDs and the re-use of the same sample ID, as the instrument appropriately detects duplicates, but not re-used sample IDs. *Duplicate sample IDs are defined as samples with the same ID concurrently being run on the instrument. *Re-used sample IDs are defined as previously processed sample IDs that are then used for a new sample. Resolution: Improvements to the detection capability for patient samples using identical barcode IDs are being identified. Share the information with laboratory staff and retain the notification as part of the laboratory Quality System documentation. If any of the affected product has been forwarded to another laboratory, provide them a copy of the letter. If you are a centralized license holder, provide the other affected laboratories of your organization or association with a copy of the letter. For assurance that the important communication has been received, respond within 10 days in one of the following ways: *Electronically, if you received the communication via email. * Manually, complete and return the enclosed Response Form. For questions regarding the notice, contact via: website: http://www.beckmancoulter.com By phone: contact your local representative.
- Initiated
- 2020-06-17
- Root cause
- Device Design
- Terminated
- 2022-08-24
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of Illinois, Ohio, Pennsylvania, and West Virginia. The countries of Australia, Canada, China, Czech Republic, France, Israel, Italy, Germany, Switzerland, Spain, Netherlands, and United Kingdom.
- Recall number
- Z-2983-2020
- Recall status
- Terminated
- CDRH record ID
- 181990
- FDA record created
- 2020-09-18
- Recall event number
- 85874