Z-2983-2020: DxA 5000; Catalog Number:B50516; UDI:15099590707118 - Product Usage: The DxA 5000 is a high-speed, modula

Recall record Z-2983-2020 for Beckman Coulter Biomedical GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2983-2020
Company in FDA record
Beckman Coulter Biomedical GmbH
FDA record creation date
Sep 18, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The DxA 5000 instructions for use may be confusing regarding the re-use of sample IDs. Beckman Coulter has received complaints where the re-use of sample barcodes were not being detected by the system .

Source facts

Action
The Urgent Medical Device Recall letters were sent via first class mail on 06/19/20. The actions to be taken by consignees are as follows: Avoid Duplicate IDs; DO NOT Re-use Sample IDs As indicated in the DxA 5000 instructions for use, ensure that only unique sample barcode IDs are used. Be aware of the difference between duplicate sample IDs and the re-use of the same sample ID, as the instrument appropriately detects duplicates, but not re-used sample IDs. *Duplicate sample IDs are defined as samples with the same ID concurrently being run on the instrument. *Re-used sample IDs are defined as previously processed sample IDs that are then used for a new sample. Resolution: Improvements to the detection capability for patient samples using identical barcode IDs are being identified. Share the information with laboratory staff and retain the notification as part of the laboratory Quality System documentation. If any of the affected product has been forwarded to another laboratory, provide them a copy of the letter. If you are a centralized license holder, provide the other affected laboratories of your organization or association with a copy of the letter. For assurance that the important communication has been received, respond within 10 days in one of the following ways: *Electronically, if you received the communication via email. * Manually, complete and return the enclosed Response Form. For questions regarding the notice, contact via: website: http://www.beckmancoulter.com By phone: contact your local representative.
Initiated
2020-06-17
Root cause
Device Design
Terminated
2022-08-24
Distribution
Worldwide distribution - US Nationwide distribution including in the states of Illinois, Ohio, Pennsylvania, and West Virginia. The countries of Australia, Canada, China, Czech Republic, France, Israel, Italy, Germany, Switzerland, Spain, Netherlands, and United Kingdom.
Recall number
Z-2983-2020
Recall status
Terminated
CDRH record ID
181990
FDA record created
2020-09-18
Recall event number
85874

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗