Z-2983-2026: Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Recall record Z-2983-2026 for Fort Defiance Industries, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2983-2026
Company in FDA record
Fort Defiance Industries, LLC
FDA classification date
Aug 18, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

Source facts

Action
Between the dates of July 8, 2026 - July 16, 2026 Fort Defiance Industries (FDI) issued a Urgent Medical Device Recall to affected consignees vi Email. FDI ask consignees to take the following actions: 1. Devices can continue in operation until the fix is implemented. 2. In the interim, users should avoid touching the WRS power connector during operation as a precaution. 3. Please contact FDI and order the upgrade kit complete with instructions to perform this Field Correction. 4. After receiving the kit, please review the instructions and determine if you are competent to perform the update. The instructions will state alternate options to complete the repair (i.e. send to military depot). 5. Perform the repair 6. FDI requests a phone call or e-mail confirmation when the upgrade is completed. This confirmation can be a phone call or e-mail message.
Initiated
2026-07-08
Root cause
Component design/selection
Distribution
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Recall number
Z-2983-2026
Recall status
Open, Classified
CDRH record ID
221436
FDA classified
2026-08-18
Recall event number
99371

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗