Open original FDA record ↗Recall record
- FDA identifier
- Z-2983-2026
- Company in FDA record
- Fort Defiance Industries, LLC
- FDA classification date
- Aug 18, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Source facts
- Action
- Between the dates of July 8, 2026 - July 16, 2026 Fort Defiance Industries (FDI) issued a Urgent Medical Device Recall to affected consignees vi Email. FDI ask consignees to take the following actions: 1. Devices can continue in operation until the fix is implemented. 2. In the interim, users should avoid touching the WRS power connector during operation as a precaution. 3. Please contact FDI and order the upgrade kit complete with instructions to perform this Field Correction. 4. After receiving the kit, please review the instructions and determine if you are competent to perform the update. The instructions will state alternate options to complete the repair (i.e. send to military depot). 5. Perform the repair 6. FDI requests a phone call or e-mail confirmation when the upgrade is completed. This confirmation can be a phone call or e-mail message.
- Initiated
- 2026-07-08
- Root cause
- Component design/selection
- Distribution
- Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
- Recall number
- Z-2983-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221436
- FDA classified
- 2026-08-18
- Recall event number
- 99371