Skip to content

Public FDA record

Z-2986-2020: Perceval Single Use Accessory Kit, Size S, REF: ICV1349, Sterile EO, Rx Only, CE, UDI: 8022057015211 - Pr

Recall record Z-2986-2020 for Sorin Group Italia S.r.l.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2986-2020
Company in FDA record
Sorin Group Italia S.r.l.
FDA record creation date
Sep 18, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.

Source facts

Action
Letter titled, Urgent Medial Device Removal, Perceval single-use accessory kit models assembly error, HV-SAL-2020-1, dated July 29, 2020 was sent to consignees via 1st class mail (certified return receipt) on July 29th, 2020 Informing the affected customers: o It could be impossible to collapse Perceval Sutureless heart valves (Perceval S and Perceval PLUS) while using specific lots of Perceval single-use accessory kits. The issue potentially affects kits assembled with a specific supplier lot; Actions to be taken by the Physician LivaNova is coordinating the removal and replacement of all potentially affected Perceval single use accessory kit lots in your inventory. Please ensure timely execution of the following actions: 1. Using the list provided in Attachment 1, please check your inventory for potentially affected devices supplied to your facility; 2. All potentially affected devices listed in Attachment 1 and still in inventory should not be used and should be quarantined, pending return of the devices to LivaNova; 3. Please complete and return the Attachment 1 by e-mail to LivaNova.FSCA@livanova.com to initiate the removal/replacement process of the Perceval single-use accessory kit(s). Your LivaNova Representative will contact you to coordinate replacement of the potentially defective device(s) if they have not already been used. Actions to be taken by the Company 1. Notification of the potentially affected devices removal via letter to inform users of the issue and advise them not to use and quarantine the potentially affected devices immediately; 2. Coordinating and providing information to the user on product replacement.Transmission of this Communication Please ensure that this notice is communicated to all personnel within your organization who need to be aware. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Please report all device-re
Initiated
2020-07-29
Root cause
Nonconforming Material/Component
Terminated
2021-05-17
Distribution
US Nationwide distribution including in the states of CA, FL, ID, ME, MN, MO, MS, NC, SD, OH, OR, PA, SD, TN, TX, VA, WA, WI and the countries of Canada, Japan, Germany, Netherland, Switzerland, Sweden, UK, Spain, Italy, France, Belgium, Finland, Greece, Slovenia, Malta, Poland, Romania, Croatia, Hungary, Serbia, Australia, Israel, Georgia, Iran, Kuwait, Oman, Quatar, Singapore, Thailand, Taiwan, Russia, South Africa.
Recall number
Z-2986-2020
Recall status
Terminated
CDRH record ID
182928
FDA record created
2020-09-18
Recall event number
86161

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗