Open original FDA record ↗Recall record
- FDA identifier
- Z-2989-2011
- Company in FDA record
- Carl Zeiss Meditec, Inc.
- FDA classification date
- Aug 10, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Carl Zeiss Meditec received a report that the INTRABEAM 20 x 20 cm sterile shield's packaging looked wet inside. Investigation found the substance to be Triethylene Glycol,
Source facts
- Action
- Carl Zeiss Meditec, Inc. sent an "Important Field Correction Action" letter dated July 13, 2011 to all affected customers. The letter included description of affected product and problem. Customers are asked to notify the firm if they have experienced any problems with the product and to return any remaining product for replacement upon completion of required testing. Customers are asked to fill out and return an Acknowledgement Form. For additional information contact Carl Zeiss Meditec at (925) 570-4844.
- Initiated
- 2011-07-01
- Root cause
- Process control
- Terminated
- 2011-10-13
- Distribution
- Nationwide (USA) Distribution including the states of FL, CA, NY, VA and AL.
- Recall number
- Z-2989-2011
- Recall status
- Terminated
- CDRH record ID
- 102395
- FDA classified
- 2011-08-10
- Recall event number
- 59362