Z-2990-2018: GE Vivid E7 ultrasound system Product Usage: Vivid E7 is a general-purpose ultrasound system, specialize

Recall record Z-2990-2018 for GE Medical Systems Ultrasound & Primary Care Diagnostics, LL. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2990-2018
Company in FDA record
GE Medical Systems Ultrasound & Primary Care Diagnostics, LL
FDA record creation date
Sep 6, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.

Source facts

Action
On March 23, 2018 GE Healthcare sent letters to their consignees informing them about the recall of LOGIQ 9, LOGIQ E9, Vivid 7, Vivid E7, Vivid E9, Vivid E80, Vivid E90, and Vivid E95 ultrasound systems which have been installed since April 2015 or which have received a power cord replacement since April 2015. Please ensure that all potential users in your facility are made aware of this safety notification and the recommended actions. Your safety and the safety of your patients is of our utmost concern. Safety Issue The system side of the power cord may break and expose the electrical conductors leading to the risk of shock. This can lead to injuries up to and including cardiac arrhythmia or cardiac arrest. A minor injury has been reported as a result of this issue. Safety Instructions You may continue to use your system. If you need to remove the power cord from the rear of the unit (see picture of location below), before doing so be sure to do the following: 1. Turn the system off 2. Unplug it from the wall If the system end of the power cord (see picture below) is damaged, discontinue use of the system until a replacement cord is provided. If you experience any power issues with your unit, discontinue use of the system and contact GEHC Service. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. Contact Information If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2018-03-22
Root cause
Component design/selection
Terminated
2021-08-24
Distribution
Nationwide and Puerto Rico FOREIGN: Algeria, Antigua and Barbuda, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State of Brazil, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Congo, The Democratic Republic of the Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, The Former Yugoslav Republic of, Malaysia, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Viet Nam
Recall number
Z-2990-2018
Recall status
Terminated
CDRH record ID
167034
FDA record created
2018-09-06
Recall event number
80928

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗