Skip to content

Public FDA record

Z-2995-2018: Portable Suction Pump, (a) Style 758010, Model CARE-E-VAC 3, Ohio Medical LLC; (b) Style 758000, Model C

Recall record Z-2995-2018 for Ohio Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2995-2018
Company in FDA record
Ohio Medical Corporation
FDA classification date
Sep 7, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The care-e-vac(R)3 Battery/AC Powered Portable Aspirator battery terminal may come in contact with the pump causing an electric short leading to damage to wiring and resulting in loss of function or an inoperable unit under battery power.

Source facts

Action
The firm, Ohio Medical, sent an "URGENT MEDICAL DEVICE CORRECTION NOTIFICATION" letter dated 8/7/2018 to its customers on 8/14/2018. The letter described the product, problem and actions to be taken. The customers were instructed to take the affected care-e-vac3 units out of service and follow the instructions, site: https://www.ohiomedical.com/4897450, provided with this notification to change the orientation of the battery; If, upon opening the device, you identify damage to wiring, please contact Ohio Medical Customer Service at 866.549.6446 or email to customer.service@ohiomedical.com for additional instructions to replace the wiring harness; and complete the attached Customer Acknowledgement Form and return it to our Quality Assurance Department at 847/855-6304 (fax), email to: CEV3_FCA@ohiomedical.com, or mail in the envelope provided to: Ohio Medical LLC, ATTN: Quality Assurance Department, 1111 Lakeside Drive, Gurnee, IL 60031. Please direct questions to Sr. Director, QA/RA at 847-855-6326 or email: barbara-ann.conway-myers@ohiomedical.com.
Initiated
2018-08-14
Root cause
Nonconforming Material/Component
Terminated
2020-12-15
Distribution
US Nationwide Distribution and internationally to countries: Canada, Saudi Arabia, and United Arab Emirates.
Recall number
Z-2995-2018
Recall status
Terminated
CDRH record ID
166447
FDA classified
2018-09-07
Recall event number
80785

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗