Open original FDA record ↗Recall record
- FDA identifier
- Z-2998-2026
- Company in FDA record
- Hollister Ulc
- FDA classification date
- Aug 21, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Hollister has identified that select lots of the catheters identified in the above table may exhibit a protrusion caused by a manufacturing defect which might lead to genitourinary trauma or bleeding.
Source facts
- Action
- Hollister issued an URGENT PRODUCT FIELD SAFETY NOTICE to its consignees on 7/16/2026 via USPS and email. The notice explained the issue, risk to health, and requested the consignee inspect each catheter for the potential presence of a protrusion (defect) located approximately 2 cm from the funnel. If a protrusion is identified, the catheter should not be used. Users should also review and follow the Instructions for Use (IFU) and ensure that the device is opened, handled, and inserted in accordance with the IFU. The notice also requested the following actions be taken: "Please follow the steps below: 1. Please complete and return the response form attached as soon as possible, even if you do not have product in your possession and send the form to ComplaintCommunication@Hollister.com. 2. If you have supplied or transferred any potentially affected product to another facility or organization, provide that facility with a copy of this letter immediately. For further information or to request replacement devices please contact us: By phone Please contact one of the following: If you are a consumer or healthcare professional, please call: 1.888.808.7456 If you are a dealer, distributor, Institution, or pharmacy, please call: 1.800.323.4060 option 1"
- Initiated
- 2026-07-14
- Root cause
- Process control
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2998-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221675
- FDA classified
- 2026-08-21
- Recall event number
- 99497