Skip to content

Public FDA record

Z-2998-2026: Onli 17cm Hydrophilic Urethral Intermittent Catheter, labeled as: 1) 82081 Ch08-17 Onli Hydrophilic Ure

Recall record Z-2998-2026 for Hollister Ulc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2998-2026
Company in FDA record
Hollister Ulc
FDA classification date
Aug 21, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Hollister has identified that select lots of the catheters identified in the above table may exhibit a protrusion caused by a manufacturing defect which might lead to genitourinary trauma or bleeding.

Source facts

Action
Hollister issued an URGENT PRODUCT FIELD SAFETY NOTICE to its consignees on 7/16/2026 via USPS and email. The notice explained the issue, risk to health, and requested the consignee inspect each catheter for the potential presence of a protrusion (defect) located approximately 2 cm from the funnel. If a protrusion is identified, the catheter should not be used. Users should also review and follow the Instructions for Use (IFU) and ensure that the device is opened, handled, and inserted in accordance with the IFU. The notice also requested the following actions be taken: "Please follow the steps below: 1. Please complete and return the response form attached as soon as possible, even if you do not have product in your possession and send the form to ComplaintCommunication@Hollister.com. 2. If you have supplied or transferred any potentially affected product to another facility or organization, provide that facility with a copy of this letter immediately. For further information or to request replacement devices please contact us: By phone Please contact one of the following: If you are a consumer or healthcare professional, please call: 1.888.808.7456 If you are a dealer, distributor, Institution, or pharmacy, please call: 1.800.323.4060 option 1"
Initiated
2026-07-14
Root cause
Process control
Distribution
US Nationwide distribution.
Recall number
Z-2998-2026
Recall status
Open, Classified
CDRH record ID
221675
FDA classified
2026-08-21
Recall event number
99497

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗