Open original FDA record ↗Recall record
- FDA identifier
- Z-2999-2020
- Company in FDA record
- ArthroCare Corporation
- FDA record creation date
- Sep 22, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.
Source facts
- Action
- On August 17,2020 the firm notified all its consignees via letter of the field action. They provided the following instructions: Please inspect your inventory and locate any devices from the listed product and batch numbers, and quarantine them immediately. " If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. If you have no product to return, please put an X in the appropriate location. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com or fax to +1-901-566-7975.
- Initiated
- 2020-08-17
- Root cause
- Mixed-up of materials/components
- Distribution
- Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT,DB,FL,GA, HI, IA, ID, IL, IN, KS, KY, MA,MD,ME,MI,MN,MO,MS,MT, NC,ND, NE, NJ, NSW, NV,NY, OH,OK, OR, PA,SC, TN,TX, UT, VA, WA, WI, WV. International Distribution: AE, AU, CH, DE, FI, FR, GB, HK, IE, PA, PR, PT, RU, SG, and ZA
- Recall number
- Z-2999-2020
- Recall status
- Open, Classified
- CDRH record ID
- 183301
- FDA record created
- 2020-09-22
- Recall event number
- 86290