Skip to content

Public FDA record

Z-2999-2020: Speed Stitch Needle Cassette

Recall record Z-2999-2020 for ArthroCare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2999-2020
Company in FDA record
ArthroCare Corporation
FDA record creation date
Sep 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.

Source facts

Action
On August 17,2020 the firm notified all its consignees via letter of the field action. They provided the following instructions: Please inspect your inventory and locate any devices from the listed product and batch numbers, and quarantine them immediately. " If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. If you have no product to return, please put an X in the appropriate location. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com or fax to +1-901-566-7975.
Initiated
2020-08-17
Root cause
Mixed-up of materials/components
Distribution
Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT,DB,FL,GA, HI, IA, ID, IL, IN, KS, KY, MA,MD,ME,MI,MN,MO,MS,MT, NC,ND, NE, NJ, NSW, NV,NY, OH,OK, OR, PA,SC, TN,TX, UT, VA, WA, WI, WV. International Distribution: AE, AU, CH, DE, FI, FR, GB, HK, IE, PA, PR, PT, RU, SG, and ZA
Recall number
Z-2999-2020
Recall status
Open, Classified
CDRH record ID
183301
FDA record created
2020-09-22
Recall event number
86290

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗