The recall was initiated because during a test on a batch of sensors on the company shipping dock, Teledyne Instruments discovered electrolyte loss from the sealing cylinder on a few sensors. This test was done pursuant to a complaint from one customer of label discoloration caused because of the electrolyte loss. The risk
of a leak form the oxygen sensor while in use is the disruption of therape
Source facts
Action
Teledyne Instruments sent emails and "Urgent--Device Field Corrections/Removal" letters in December 2010 to all affected customers.
Th letter includes description of device and problem. Customers are advised to check them before use for any signs of electrolyte loss and if discoloration is noted to discontinue its use. and provides recommendation. Teledyne-desplaines to generate a product field corrections/removal log that lists names/addresses of the consignees, quantity of sensors shipped, sensor serial numbers and the date of shipment.
For additional information or questions contact Teledyne Instruments at (626) 934-1504.
Initiated
2010-11-29
Root cause
Process design
Terminated
2011-08-22
Distribution
Worldwide Distribution: Nationwide (US) distribution including the states of CA, TN and WI; and the countries of , Canada, Brazil, Chile, Saudi Arabia, and the U.K.
Recall number
Z-3001-2011
Recall status
Terminated
CDRH record ID
98984
FDA classified
2011-08-11
Recall event number
58322