Z-3003-2026: American Contract System Convenience Kit, D & C PACK, Model/Catalog Number: LLPD90H

Recall record Z-3003-2026 for American Contract Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3003-2026
Company in FDA record
American Contract Systems Inc
FDA classification date
Aug 21, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Convenience Kits contain Medline's Red Rubber Intermittent Catheters (8Fr-22 Fr) and Red Rubber Intermittent Catheter Trays. Medline Industries, LP has identified representative samples did not meet applicable performance specifications.

Source facts

Action
An URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter dated 6/29/26 was sent to customers. ACTIONS TO TAKE: "Identify, segregate, and quarantine all impacted product. Other components within the convenience kits can still be used once kit is properly labeled with the warning label below. "Add warning labels to affected ACS kits to ensure all users are aware of the actions below and identified in the enclosed Medline recall notice. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. "Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have, or to request printed labels. "Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Complete the attached Recall Response Form and return by July 17, 2026. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com . Your assistance is appreciated and necessary to prevent any risk to patient or medical professional health as a result of not following these instructions.
Initiated
2026-06-29
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of IA, OH.
Recall number
Z-3003-2026
Recall status
Open, Classified
CDRH record ID
221726
FDA classified
2026-08-21
Recall event number
99514

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗