Open original FDA record ↗Recall record
- FDA identifier
- Z-3004-2026
- Company in FDA record
- Merge Healthcare, Inc.
- FDA classification date
- Aug 21, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software issue may cause measurements to be displayed inaccurately.
Source facts
- Action
- Merge Healthcare issued an Urgent Field Safety Notice to its consignees on 7/22/2026 via email with a follow-up telephone call. The notice explained the issue and how to identify it, risk to patient, and requested the following: "Immediate actions to reduce risk to patients Merge Healthcare recommends that customers contact Merge Healthcare Support to identify any affected ultrasound studies and the necessary next steps to be taken. Until the software update is installed, ultrasound measurements performed using the Merge PACS & OrthoPACS Viewer should be confirmed against the measurements acquired at the ultrasound modality." The required actions are as follows: "Merge Healthcare Support will contact all customers to: i) Confirm that you have read and understood this Urgent Field Safety Notice. ii) Assist you in identifying and reviewing any ultrasound studies with measurements that may have been affected. iii) Schedule the installation of the software update that eliminates the root cause of this issue. Please forward this notice to any individuals within the organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge.
- Initiated
- 2026-07-22
- Root cause
- Software Design Change
- Distribution
- US Nationwide distribution in the states of California, Florida, Kentucky, Nebraska, Washington.
- Recall number
- Z-3004-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221734
- FDA classified
- 2026-08-21
- Recall event number
- 99520